Results of the first randomized French study evaluating self-testing of the International Normalized Ratio.

Pas encore traduit Pas encore traduit
Catégorie Primary study
JournalThe Journal of heart valve disease
Year 2011
Loading references information

BACKGROUND AND AIM OF THE STUDY:

In this single-center French study, the conventional management of patients receiving vitamin K antagonists (VKAs) was compared with an International Normalized Ratio (INR) self-testing program. The aim was to determine the reliability of self-testing devices, and to estimate the variability of the self-measured INR within the therapeutic and target range.

METHODS:

A total of 206 patients who had undergone valve replacement with a mechanical prosthesis, with or without myocardial revascularization, between May 2004 and September 2007 was randomized into two groups. Group 1 patients (n = 103) underwent INR monitoring at a laboratory, while Group 2 patients (n = 103) underwent self-testing INR using either the CoaguChek (Roche) (Group 2A; n = 55) or INRatio (Hemosense) (Group 2B; n = 48) system. Patients in Group 1 underwent at least once-monthly INR measurement, while those in Group 2 carried out once-weekly self-testing, and also underwent once-monthly INR measurement at the laboratory. The large majority of patients (97.9%) were treated with fluindione.

RESULTS:

The mean follow up period was 49.0 +/- 10.3 weeks. Self-testing was reliable, with a correlation coefficient between device- and laboratory-measured INRs of 0.80 [CI: 0.78, 0.82] (p < 0.0001). The proportion of time spent within the INR target range was significantly higher for Group 2 (61.5 +/- 19.3% versus 55.5 +/- 19.9%; p < 0.05), while the absolute mean deviation of INR from the target range was higher in Group 1 (60.1 +/- 70.2% versus 47.4 +/- 51.5%). Adverse events were reported by seven patients in Group 1, but by no patients in Group 2 (p < 0.01).

CONCLUSION:

INR self-testing devices are reliable and beneficial. Moreover, INR self-monitoring allows an enhanced stability within a target range, and also helps to prevent serious postoperative complications.